intelligenthasem.blogg.se

Fm8 review
Fm8 review








fm8 review

The “inputs” are the test fixture irradiance (above) and test diffuser BRDF.Radiance calibration “output” is instrument response to a known radiance target.See backup slides for lamp and mirror trend data.Redundancy in sources and test fixtures ensure irradiance calibration accuracy.The “inputs” are lamp irradiance and collimating mirror reflectance.Irradiance calibration “output” is instrument response to a known source of irradiance (the irradiance of our test fixture).Radiometric Accuracy for SBUV/2 Instruments Radiometric Accuracy Requirements from GSFC S-480-31C No requirement for radiometric accuracy below 200 nm.

fm8 review

Radiance errors assume test diffuser calibration accuracy of.We use deuterium lamps in this wavelength range.6% for argon mini-arc lamps 200-250 nm.Irradiance errors assume lamp calibration accuracy of.Absolute radiometric accuracy for the instrument as shown in the charts below (on the following slide).“Quartz iodide, argon arc or other standard sources of spectral irradiance used for preflight calibration shall be the best sources that can be provided by the National Bureau of Standards”.The GSFC specification for SBUV/2 requires:.Actual meeting agenda will be presented on-screen.Radiometric calibration constants for FM8.Comparison of radiometric calibration in 20.

fm8 review

  • Radiance calibrations on USSF and NITF.
  • Normalization of vacuum irradiance calibration.
  • Comparison of calibration constants on the PTF in 20.
  • Introduction to FM8 radiometric calibration.
  • SBUV/2 Flight Model 8Radiometric Calibration Data Review Kevin Kelly Zongying Wei Test Engineer Optical Engineer (303) 939-5839 (303) 939-6898










    Fm8 review